Introduction, Introduction 13, 20 pressure intensifier introduction – MTS Model 286-20 Confining Pressure Intensifier User Manual

Page 13: Fluid resivoir control panel

Advertising
background image

286.20 Pressure Intensifier

Introduction

13

Introduction

This section describes the Model 286.20 Confining Pressure Intensifier (CPI), its
role in the typical rock mechanics test system and other system components
typically used with the Confining Pressure Intensifier. Specifications and
dimensions are also included.

The Model 286.20 Confining Pressure Intensifier is designed to provide a source
of fluid at pressures suitable for use in the confining pressure chamber of a
triaxial cell. The CPI is used to fill the triaxial cell with confining fluid,
pressurize the fluid, control the pressure, and empty the fluid from the triaxial
cell at the conclusion of a test.

Some model versions produce confining pressures up to 140 MPa (20,000 psi).
When connected to a fluid-filled triaxial cell, the maximum fluid volume that can
flow during pressurization of the external device and its specimen can be as high
as 1300 cubic centimeters (80 cubic inches). See

“Specifications and

Dimensions”

on page 18 for specifications for all models.

Triaxial

Cell

Open

Close

Confining Fluid Input

Open

Close

Confining Fluid Output

Open

Close

Reservoir Fill

Fluid

Reservoir

Open

Close

Intensifier Fill

Pressure

Intensifier

Low Pressure

Relief

Pressure

Transducer

Low pressure line

High pressure line

286.20

Confining

Pressure

Intensifier

r

Fluid Filter

Off

Fluid Transfer Pump Control

Air supply in

100 psi max.

Air Driven

Hydraulic

Pump

Fluid Direction

A

B

C

C

A

B

Empty

Reservoir

From fluid

supply or to

empty fluid

Fill system from

fluid supply

using pump

Fill system

from reservoir

using pump

Confining

Pressure

A

B

A

B

Air Direction

Air supply into empty

traxial cell.

Bleed air into reservoir

when filling triaxial cell.

psi

bar

High Pressure

Relief

Air Pressure

Relief

815

Fluid
Resivoir

Control
Panel