20 pressure intensifier, Operating procedures operation – MTS Model 286-20 Confining Pressure Intensifier User Manual
Page 40
![background image](/manuals/568023/40/background.png)
286.20 Pressure Intensifier
40
Operating Procedures
Operation
Triaxial
Cell
Open
Close
Confining Fluid Input
Open
Close
Confining Fluid Output
Open
Close
Reservoir Fill
Fluid
Reservoir
Open
Close
Intensifier Fill
Pressure
Intensifier
Low Pressure
Relief
Pressure
Transducer
Low pressure line
High pressure line
286.20
Confining
Pressure
Intensifier
r
Fluid Filter
Off
Fluid Transfer Pump Control
Air supply in
100 psi max.
Air Driven
Hydraulic
Pump
Fluid Direction
A
B
C
C
A
B
Empty
Reservoir
From fluid
supply or to
empty fluid
Fill system from
fluid supply
using pump
Fill system
from reservoir
using pump
Confining
Pressure
A
B
A
B
Air Direction
Air supply into empty
traxial cell.
Bleed air into reservoir
when filling triaxial cell.
psi
bar
High Pressure
Relief
Air Pressure
Relief
3 - Open
7 - Close
4 - Close
10 - Open
16 - Close
Pressurizing and Depressurizing The Confining Fluid
18 - Open
2 - Close the
confining
pressure
valve
9 - Open the
confining
pressure
valve