0 run the test, 0 temperature parameters, Run the test – MTS Fatigue Template User Manual

Page 106

Advertising
background image

Description

Parameter

Specify the length on the specimen that has the
reduced cross-sectional area.

Extensometer Gage
Length (L)

Extensometer Gage Length is a specimen variable
that other variables may use in a calculation.

The value you specify here represents the slope
value of the material’s stress-strain curve in the
elastic deformation region.

Elastic Modulus (E)

Elastic Modulus is a specimen variable that other
variables may use in a calculation.

Specify the stress value at which permanent
deformation occurs.

Yield Strength

Yield Strength is a specimen variable that other
variables may use in a calculation.

Specify the temperature at which Elastic Modulus
and Yield Strength are valid.

Valid at Temperature

11.2.0 Run the Test

11.2.1.0 Temperature Parameters

Temperature parameters for the low-cycle fatigue strain test include:

High Temperature Difference

Low Temperature Difference

Soak Time

Start Temperature

Start Temperature Limit

Temperature Ramp

Time to Start Temperature

Zone 1 Temperature

Zone 2 Temperature

Zone 3 Temperature

The application checks the values you enter to make sure they are within the system capabilities.

106 MTS TestSuite

11.0 ASTM Trapezoidal Strain Elevated Temperature Template

Advertising