8.2.3.0 Data Storage Parameters

Data Storage Parameters Properties

Description

Item

Sets the number of cycles to save to disk at the end
of the test or when the test stops.

Ending Cycles

Sets the amount of change that must occur in the
maximum load between saves to disk.

Load Storage
Change

Sets the value of N, which is the rate (number) of
peak-valley cycles for a store to disk to occur. For

PV Nth Cycle
Stored

example, if this parameter is 100, data is saved to
disk every 100 cycles.

Sets the number of cycles to save to disk at the
beginning of the test or when the test is started.

Starting Cycles

Sets the number of cycles of time data per decade
based on a logarithm to save to disk.

Time Cycles Per
Log Decade
Stored

Sets the number of levels between saves to disk.

Number of Levels

Sets the range the data must exceed to detect the
peak-valley data. Within this range, points are
considered noise and not peak or valley data.

Strain Noise Band
Percent

Sets the amount of change that must occur in the
displacement between saves to disk.

Displacement
Storage Change

8.2.4.0 Load Termination Parameters

Load Termination Parameters

Description

Item

Specify a number of cycles for the maximum length
of the test.

Total Cycles

Specify a number of cycles after which the specimen
is expected to be stable and beyond hardening or
softening.

Cycles for Stable
Cycle

Specify a percentage used to calculate stability.

Stable Cycle
Percent

MTS TestSuite 69