0 set up the test, 0 astm low-cycle fatigue (lcf) strain template, 0 set up a test – MTS Fatigue Template User Manual

Page 40: Set up the test

Advertising
background image

6.1.0 Set Up the Test

6.1.1.0 ASTM Low-Cycle Fatigue (LCF) Strain Template

The Low-Cycle Strain template applies a small number of high-amplitude cyclic forces to the specimen to
determine the maximum load or fatigue life of a material. A variation of the template performs the test at an
elevated temperature. Both templates are designed to be in compliance with ASTM standard E606-04 and
D3479-07.

After a test run and a specimen geometry are added, a window provides access to all test parameters and
test operation steps. The buttons enable you to perform the steps to run a low-cycle fatigue test. The general
order of the buttons is left to right, from the top row to the bottom row.

As the test runs, you can stop the test to change test parameters. After the test is complete, the data can be
analyzed.

The main features of the low-cycle fatigue template include:

Buttons enable you to:

Set up all test parameters

Verify and reattach the extensometer

Measure the modulus

Run the test

Create a report

End the test

Comprehensive monitor views help you monitor test progress.

Tabular and graphical displays help you review test results.

Analysis definitions provide analysis results.

Reports can be generated after the analysis is defined.

6.1.2.0 Set Up a Test

To create and run a new test:

1. Create a new test from a template.

a) Click File > New > Test from Template.
b) In the Create from Existing Test window, select a test.
c) Click OK.

The new test is automatically created and assigned a default name. You can change the name and
enter comments about the new test by clicking the Edit button and making the changes.

2. Create a new test run.

a) Click New Test Run button.

40 MTS TestSuite

6.0 ASTM Low-Cycle Fatigue (LCF) Elevated Temperature Template