MTS Axial Extensometers User Manual

Page 28

Advertising
background image

Configuration

28

10. Place one edge of a machinist’s square against the front edge of the

top extensometer arm. Adjust the attachment adapter until the
alignment rod contacts the perpendicular edge of the square.

.

11. To align the alignment rods vertically, place the flat surface of the

machinist’s square against the alignment rod in the upper
attachment adapter. Adjust the lower attachment adapter until both
alignment rods are resting against the surface of the square.

Attachment Adapter Alignment

12. If straight knife edges are being used, place the flat surface of the

machinist’s square across the face of both knife edges to ensure
that both edges are parallel to each other.

13. Being careful not to change the attachment adapter alignment, use

a small adjustable wrench to grasp the attachment adapter.

Perpendicular

Alignment of

Rod and Square

90o

Slider

Spring

Attachment

Adapter

Square

Square with the flat

surface against the

upper alignment rod.

Alignment

Rod

Vertical Alignment of

Alignment Rod

Adjust the lower

attachment adapter until

the alignment rod rests

against the square.

Advertising