How to use gage length de-extenders, How to use gage length de-extenders 37 – MTS Axial Extensometers User Manual

Page 37

Advertising
background image

Configuration

37

How to Use Gage Length De-Extenders

Gage length de-extenders let you change where the extensometer
contacts the specimen. The following figure shows a typical gage length
de-extender.

The following procedure describes a typical gage length de-extender
installation. Actual installation may vary. See the installation drawing
accompanying the gage length de-extender for specific details.

1. Zero the arms of the extensometer. Depending on the extensometer,

this can be accomplished using the zero pin, stop block, or a special
fixture.

2. Remove the existing knife edge hold-down or quick attachment

adapter from the upper arm of the extensometer. Take care not to
loosen the spring attachment hooks and knife edges.

Note

A knife edge hold-down and spring attachment hook for small
diameter specimens are already attached to the de-extender.

3. Loosely attach a gage length de-extender to one arm of the

extensometer. Use the socket head cap screws which secured the
fixtures removed in Step 2.

4. While pressing the de-extender against the end of the extensometer

arm, tighten the two socket head cap screws.

5. Repeat Step 3 and Step 4 for the other extensometer arm.

6. Follow the appropriate procedure and recalibrate the extensometer.

m

MODEL

xxx.xxx-xx

Each gage length de-extender
comes with a knife edge hold-
down, knife edge and spring
attachment hook already
installed.

De-extender

Knife Edge

Cap Screw

Hold-down