Specifications, Specifications 17 – MTS Series LX Laser Extensometer User Manual

Page 17

Advertising
background image

Specifications

LX Laser Extensometer

Introduction

17

Specifications

W

H

L

P

ARAMETER

T

YPICAL

S

PECIFICATION*

Dimensions
(Length [L] x Width [W] x Height [H])

LX 500

LX 1500

381 x 114 x 180 mm (15 x 4.5 x 7.1 in)

686 x 196 x 442 mm (27 x 7.7 x 17.4 in)

Weight

LX 500

LX 1500

5.9 kg (13 lb)

21.8 kg (48 lb)

Power Input

Voltage

Current

Frequency

115 V AC (standard), or 230 V AC (optional), ± 10%

120 mA @ 115 V AC, 60 mA @ 230 V AC

50/60 Hz

Operating Conditions

18 ºC to 28 ºC (64 ºF to 82 ºF)
25% to 75% relative humidity, noncondensing
0 to 2745 m (0 to 9000 ft) altitude

Mounting Provisions

4 holes (1/4 - 20 thread) located in base

Laser Source

670 nm Diode Laser
<1 mW maximum scanned output
CDRH compliant Class II Laser Instrument

Advertising