MTS Series LX Laser Extensometer User Manual

Page 18

Advertising
background image

LX Laser Extensometer

Specifications

Introduction

18

Scan Line Orientation

Vertical

Scan Rate

100 scans/second

Unit System

Selectable, millimeter or inch

Target Distance

LX 500

LX 1500

Selectable: 304 mm (12 in), 381 mm (15 in)

Selectable: 381 mm (15 in), 457 mm (18 in)

Measurement Range

LX 500

LX 1500

5 to 127 mm (0.2 to 5.0 in)

8 to 381 mm (0.3 to 15 in)

Resolution

LX 500

LX 1500

0.001 mm (0.0001 in)

0.01 mm (0.001 in)

Nonlinearity (maximum)

LX 500

LX 1500

±0.025 mm (±0.0001 in)

±0.15 mm (±0.006 in)

Repeatability

LX 500

LX 1500

±0.003 mm (±0.0001 in)

±0.03 mm (±0.001 in)

Zero Suppression

Reading can be zeroed anywhere within the full scale range of the
instrument

Multiple Scan Averaging

A moving window average is applied over a programmable number
of scans to digitally filter the output readings

Display

2-line, 16 characters/line LCD type display with LED backlighting

Analog Output Range

±10 V full scale (±5 V full scale optional)

Serial I/0 Port

Standard 3-wire type, RS-232
Selectable baud (9600, 4800, 2400, 1200)

Options

Consult MTS

*

Specifications apply at 25

°

C (77

°

F), 50% relative humidity.

† Contact factory for other ranges.

‡ Over optimum displacement range at calibrated target distance.

P

ARAMETER

T

YPICAL

S

PECIFICATION*

Advertising