Target Placement

LX Laser Extensometer

Operation

37

Target Placement

Depending on the measurement mode selected, the laser extensometer measures
the distance between the front edges of targets T

1

and T

2

, or the distance between

the front edges of targets T

1

and T

2

, T

2

and T

3

, T

3

and T

4

successively during

each scan.

The target edges are labeled as follows:

T

1

is the front edge of target 1

T

2

is the front edge of target 2

T

3

is the front edge of target 3

T

4

is the front edge of target 4

In the T

2

- T

l

measurement mode, the targets T

1

and T

2

are positioned such that

the front edges are coincident with the gauge length. In the multisegment
measurement mode, the targets T

1

, T

2

, T

3

, and T

4

are arranged sequentially with

the front edges of T

1

and T

4

aligned with the gauge length.