Shunt error status, Shunt error status 58, Refer to – MTS SWIFT 40 Sensor User Manual
Page 58: Shunt, Error status
Advertising
![background image](/manuals/568057/58/background.png)
SWIFT 40 Sensors
58
Shunt Error Status
Software Utilities
Shunt Error Status
This is the error status for each bridge from the last shunt cal that was performed.
The error status is a bit-mapped status word, where the error is present if the bit is
set to 1.
Bit 0 = Bridge amplifier was saturated prior to applying the shunt.
Bit 1 = Bridge amplifier was saturated after applying the shunt.
Bit 2 = Measured delta value failed high.
Bit 3 = Measured delta value failed low.
Bit 4 = A procedural error occurred (unable to read a value, etc.).
In the following example the error status received is 5 indicating that bits 0 and 2
were set.
Bit
4
3
2
1
0
Bit Value
16
8
4
2
1
Example
0
0
1
0
1
Error Status=
5