User interface, Setup settings, Leds indicate the status of the heater – New Era HEATER-KIT-1LG Syringe Heater User Manual

Page 4: Keypad

Advertising
background image

New Era Pump Systems Inc.

www.SyringeHeater.com

Model HEATER-KIT-1

Publication #1200-01

3

04/23/14

SETUP

ENTER

STOP

ACTIVE

ACTIVE

OUT

ALARM

C

O

M

P

U

T

E

R

P

U

M

P

4.

User Interface

The controller will display, as the default display, the current
temperature of the primary heating pad, in the currently selected units,
C or F.

LEDs indicate the status of the heater.

ACTIVE

Temperature Set Point is actively being
maintained

OUT

Heating pads are being powered. Flashing
indicates on/off power duty cycle to slow
down heating when near the Set Point

ALARM

Heating pad temperature is greater than 20
degrees C above the Set Point

Keypad

Up / Down Arrow Buttons

Increases or decreases settings. By default, will change the
temperature Set Point
Settings with decimal points: Press both buttons simulaneously to
shift the decimal point.

SETUP Button

Displays the setup menu. Each press will display the next menu
selection.

ENTER / STOP Button

When displaying the setup menu or changing a value, will select
the entry or save the new value. The display will blink when the
new value is saved in non-volatile memory.

Otherwise, will stop the heater. The Active LED will turn off.
Temperature will no longer be maintained.

Active Mode Buttons

Simultaneously press the SETUP and ENTER/STOP buttons. The
Active LED will turn on and the heatings pads will start heating
until the temperature reaches the Set Point.

5.

Setup Settings

Press the SETUP button

to display the setup menu. Each press of the Setup button will display

the next menu entry. All settings are saved in non-volatile memory. Pressing an Up or Down arrow button
will immediately select Temperature Clamping Set Point and will increment or decrement the setting.

Press ENTER

to select a menu entry or accept and enter a setting, exiting Setup Configuration.

A value being changed will be automatically entered after a few seconds time out period. The display will