New Era HEATER-KIT-1LG Syringe Heater User Manual

Page 5

Advertising
background image

New Era Pump Systems Inc.

www.SyringeHeater.com

Model HEATER-KIT-1

Publication #1200-01

4

04/23/14

Press an Up or Down arrow button

or

to change a displayed setting.


The setup menu is as follows:

Indicates that the display alternates between 2 displays.

Temperature Clamp Set Point

When active, the heater will heat the pads until the set point is
reached on the primary pad. Use the Fiine Tune settings to
control the regulation of the temperature clamping.

Alarm Set Point will be 20 degrees C (36 F) above the Set Point.

Fine Tune Temperature Set Point

Fine Tunes thermo-kinetic settings to improve regulation of
temperature Set Point clamping.

Press

to display the Fine Tune settings sub-menu.

Slow Down Temperature Delta

Degrees below Set Point to begin heating slow down.

Heating duty cycle will be reduced based on degrees below
temperature set point and rate of temperature change.
Set to 0 to disable. Default = 10.

Set Point heat clamping duty cycle
base percentage (Pc).

Hold duty cycle percentage is the base heat duty cycle percentage
to trickle charge the heaters to clamp the temperature setting.
Actual percentage will be dynamically adjusted up or down
according to temperature feedback
Set to 0 to disable. Default = 10.

This setting will not appear in Active Mode. C is default setting

Set temperature units: C or F (Celsius or Farenheit). When the
units are changed, all settings will change units without being
converted.

n

Enable Lock Out Mode. Set to 1 to enable. 0 is default setting.

Press

when enabled to set Lock Out code.

nnnn

Lock Out code: use arrow keys to set blinking digit.

Press

or

to move to next digit.

n

Power failure mode. Set to 1 to enable. 0 is default setting.

When Power Failure Mode is enabled, if the heater loses power

while in Active Mode, the heater will power up in Active Mode
when power is restored. Alarm mode will override Active Mode.

nn

RS-232 communications network address. 0 is default setting.

Sets the RS-232 network address for use with a network of
heaters and pumps connected to a computer. 00 is the default
address.

ENTER

STOP

SETUP

ENTER

STOP

ENTER

STOP