Lock out mode – New Era HEATER-KIT-1LG Syringe Heater User Manual

Page 8

Advertising
background image

New Era Pump Systems Inc.

www.SyringeHeater.com

Model HEATER-KIT-1

Publication #1200-01

7

04/23/14

4)

Finalize the heating pad calibration.

The calibration entry process can be repeated before the process is finalized, or the entries can be

ignored if the process is not finalized. Turning off power will discard the entries.

To finalize the calibration, enter the command:

CAL

If the calibration data was valid, the heater will respond with:

OK

If there was an error in the calibration, the heater will respond with:

?NA

After calibration, the heating pad number will be set to 0. Selecting pad number 1, will change the
calibration back to the factory defaults:

PAD 1

6.4

Lock Out Mode

When enabled, Lock Out mode will lock out the keypad until the user defined lock out code is entered.
If any key is pressed while locked out, the display will show:

Pressing either arrow key will enter Lock Out code entry mode:

nnnn

. Any other key will return to

the previous display.

Use the arrow keys to set the blinking digit. Press the Setup or Enter keys to move to the next digit.

If the code entered is wrong, the display will show:

. Press any key to clear.

If the code is correct, the keypad will be unlocked, but Lock Out setting will remain enabled. The next
time power is turned out, the keypad will be locked out. Use Setup mode to disable Lock Out mode.

6.4.1

Default Code Bypass

If the lock out code is forgotten, the default code can be entered: While turning on power to the unit, press
and hold the Setup key. Lock Out code entry mode will be displayed. To unlock the keypad, enter the

code:

1392

.