Pre-defined program selection, Example – New Era FW-1-X FIRMWARE UPGRADE NE-1000X FAMILY OF SYRINGE PUMPS User Manual

Page 4

Advertising
background image

NE-1000X

4

www.SyringePump.com

ET:RS

Reset all Event traps

RET

End Sub-Routine and return to Phase after calling Phase number

BEEP

Sound a single beep

n

OUT.0

Set program output logic pin low or high

CLR.D

Clear total volume dispensed.

tr:Ft

Temporary trigger input configuration. (“Ft” is configuration sub-menu)

3.

Pre-Defined Program Selection

Press and hold the "Program Function" (Volume) key while turning on power to the pump. This will pre-load the Continuous
Pumping Program. See below for details.

In addition, this will set the pump’s RS-232 communication to Continuous pump control with a second NE-1000X syringe pump.
Cable CBL-DUAL-3 is required to control the second pump.

4.

Example

Linear Ramp, Sub-Routines, Increment, Decrement

Rate 100 ml/hr

Decrease 100 ml/hr

Sub-Routine Calls
to Linear Ramp

Rate 500 ml/hr

Increase 200 ml/hr

Repeat Program
Rate 100 ml/hr


Phase

Function

Rate

Volume

Direction

1

RATE

100 ml/hr

1.0 ml

Infuse

Phase

Function

Rate

Time

Direction

2

INCR

200

00:10

Infuse

Phase

Function

3

JP:08

Phase

Function

Rate

Volume

Direction

4

RATE

500 ml/hr

5.0 ml

Infuse

Phase

Function

Rate

Time

Direction

5

DECR

100

00:10

Infuse

Phase

Function

6

JP:08

Phase

Function

7

JP:01

Increase rate 200 ml/hr and
continue infusing for 10 seconds

Start Infusing at 100 ml/hr for 1 ml

Call Linear Ramp Sub-Routine

Decrease rate 100 ml/hr and
continue infusing for 10 seconds

Infuse at 500 ml/hr for 5 ml

Call Linear Ramp Sub-Routine

Repeat Program