Dual pump gradient pumping mode, Pump 1 + pump 2 = constant rate, Requirement – New Era FW-1-X FIRMWARE UPGRADE NE-1000X FAMILY OF SYRINGE PUMPS User Manual

Page 8: Quick setup, Setup

Advertising
background image

NE-1000X

8

www.SyringePump.com

Dual Pump Gradient Pumping Mode












Maintains a constant total flow rate between 2 syringe pumps. A gradient (linear function)
dispensing function is programmed into the Master pump. The secondary pump will be
automatically programmed with the inverse of the Master pump’s gradient function. The sum
of the two pump’s flow rates will be a constant.

Requirement

:

1)

2 Pumps from the NE-1000 syringe pump family with ‘X firmware upgrades.

2)

CBL-DUAL-3 dual pump communications cable.

Quick Setup:

(On the Master pump only – DO NOT change default settings on the second pump)

1)

Attach the communications cable, CBL-DUAL-3, to the “To Computer” port on the back of both

pumps.

2)

Select the default Gradient Mode pumping program:

- Turn on power to the pump WHILE pressing the “Volume / Program Function” key.
- Use any up-arrow key to select Gradient Mode, displayed as “Grad”.
- Press the “Volume” key to select, or wait for the time out.

3)

Set the syringe diameter on the Master pump. Both pumps will use the same diameter.

(Default program assumes at least a 14.43 mm diameter).

4)

Press “Start” on the Master pump to start the dual pump Gradient pumping program.

Setup:

(On the Master pump only – DO NOT change default settings on the second pump)

1)

Attach the dual pump cable, CBL-DUAL-3, to the “To Computer” port on both pumps.

2)

Select Gradient communications mode:

From the Setup menu (Press and hold “Diameter / Setup” key), select Gradient communications
Mode (displayed as Grad). When Ad:00 (or other communications mode) is displayed, use the
left-most arrow key to select.

3)

Set the syringe diameter.

P

u

m

p

in

g

R

a

te

Time

Pump 1 + Pump 2 = Constant Rate