New Era FW-1-X FIRMWARE UPGRADE NE-1000X FAMILY OF SYRINGE PUMPS User Manual
Page 5
 
NE-1000X
5
www.SyringePump.com
 
 
 
 
 
 
 
 
Phase
Function
Rate
Time
Direction
8
LINR
0 ml/hr
00:10
Infuse
Phase
Function
Rate
Time
Direction
9
LINR
1000 ml/hr
00:00
Infuse
Phase
Function
Rate
Time
Direction
10
LINR
1000 ml/hr
00:10
Infuse
Phase
Function
Rate
Time
Direction
11
LINR
0 ml/hr
00:00
Infuse
Phase
Function
12
RET
 
 
 
The above program would be entered from RS-232 with the following commands: 
PHN 1 
FUN RAT 
RAT 100 MH 
VOL 1 
DIR INF 
PHN 2 
FUN INC 
RAT 200 
TIM 00:10 
DIR INF 
PHN 3 
FUN JMP 08 
PHN 4 
FUN RAT 
RAT 500 MH 
VOL 5.0 
DIR INF 
PHN 5 
FUN DEC 
RAT 100.0 
TIM 00:10 
DIR INF 
PHN 6 
FUN JMP 08 
PHN 7 
FUN JMP 01 
PHN 8 
FUN LIN 
RAT 0.0 
TIM 00:10 
DIR INF 
PHN 9 
FUN LIN 
RAT 1000 MH 
TIM 00:00 
DIR INF 
PHN 10 
FUN LIN 
RAT 1000 MH 
TIM 00:10 
DIR INF 
PHN 11 
FUN LIN 
RAT 0.0 MH 
TIM 00:00 
DIR INF 
PHN 12 
FUN RET 
 
 
 
 
 
 
 
 
 
 
 
 
 
Return to Phase following calling 
Program function 
 
Linearly ramp the infusion rate 
from 1000 ml/hr to 0 ml/hr over 10 
minutes 
 
Linearly ramp the infusion rate 
from 0 ml/hr to 1000 ml/hr over 10 
minutes 
 
Linear Ramping Sub-Routine