New Era FW-1-X FIRMWARE UPGRADE NE-1000X FAMILY OF SYRINGE PUMPS User Manual

Page 7

Advertising
background image

NE-1000X

7

www.SyringePump.com

- Master pump starts infusing
- Second pump primes (if enabled), then pauses

- When the infusing pump reaches the overlap position, if enabled, or changes pumping direction, the paused

pump will start to infuse

- Sequence repeats:

- Refilling pump fills, primes, pausing and waits for the other pump.
- Infusing pump infuses to the overlap position or syringe empty, and then signals the other pump to start

infusing. When empty, changes direction and starts refilling

While Pumping
The pumping rate can be changed. The new pumping rate will be transmitted to the other pump, into the
corresponding pumping sequence.

Pressing the Stop key on either pump, will Pause the sequence on both pumps. Pressing the Start key will
continue pumping.

Notes
The refill rate needs to be fast enough so that the syringe is refilled and primed before the infusing pump has
emptied.

If the refilling pump is still pumping when the infusing pump empties, an error message will be displayed and the
pumps will stop.

Default Continuous Pumping Program
To load the default program: On the Master pump, while turning on power:
Press and hold the “Volume”/”Program Function” key until [Cont] is displayed. Press the “Volume” button again
to enter the selection.

Default Continuous Mode Pumping Parameters

Phase #

Function

Rate

Volume

1

RATE

Infusion

100.0 mL/hr

Syringe

50.00 mL

2

RATE

Refill

500.0 mL/hr

Overlap

0.000 mL

3

RATE

Prime

200.0 mL/hr

Prime

0.100 mL

Communications Mode set to Continuous: [Cont]