- Master pump starts infusing 
- Second pump primes (if enabled), then pauses 
- When the infusing pump reaches the overlap position, if enabled, or changes pumping direction, the paused
pump will start to infuse
- Sequence repeats:
- Refilling pump fills, primes, pausing and waits for the other pump. 
- Infusing pump infuses to the overlap position or syringe empty, and then signals the other pump to start 
infusing. When empty, changes direction and starts refilling
While Pumping 
The pumping rate can be changed. The new pumping rate will be transmitted to the other pump, into the 
corresponding pumping sequence. 
 
Pressing the Stop key on either pump, will Pause the sequence on both pumps. Pressing the Start key will 
continue pumping. 
 
Notes 
The refill rate needs to be fast enough so that the syringe is refilled and primed before the infusing pump has 
emptied. 
 
If the refilling pump is still pumping when the infusing pump empties, an error message will be displayed and the 
pumps will stop. 
 
Default Continuous Pumping Program 
To load the default program: On the Master pump, while turning on power: 
Press and hold the “Volume”/”Program Function” key until [Cont] is displayed. Press the “Volume” button again 
to enter the selection. 
 
Default Continuous Mode Pumping Parameters
Phase #
Function
Rate
Volume
1
RATE
Infusion
100.0 mL/hr
Syringe
50.00 mL
2
RATE
Refill
500.0 mL/hr
Overlap
0.000 mL
3
RATE
Prime
200.0 mL/hr
Prime
0.100 mL
Communications Mode set to Continuous: [Cont]