Safety regulations for the as200, Extract from the individual chapters – Retsch AS 200 User Manual

Page 30

Advertising
background image

31.07.2012

Retsch GmbH

30

Safety regulations for the AS200

Extract from the individual chapters

Subject

Action

Hazards

Vibration height

Vibration height >2.5 mm in the
AS200 basic and digit models

Armature strikes the magnets;
mechanical components could
be damaged.

Shipping

During shipment the AS200 is not
to be subjected to impact or
vibration; do not throw the
package.

Electronic

and

mechanical

components could be damaged.

Temperature
fluctuation

In

case

of

temperature

fluctuations, protect the AS200
against condensation.

Electronic components could be
damaged.

Scope of supply

If the shipment is not complete or
is damaged, you must notify the
forwarder and Retsch GmbH
immediately (within 24 hours).

Under certain circumstances
compensation cannot be paid
for claims which are lodged
later.

Ambient
temperature

Temperature below 5°C
Temperature above 40°C

Electronic components could be
damaged.

Humidity

Exceeding

80%

R.H.

at

temperatures up to 31°C

Electronic components could be
damaged.

Installation

Shipping nuts not removed.

Mechanical

and

electronic

components could be damaged.

Electrical
connection

Local power supply does not
correspond to the values given on
the data plate.

Mechanical

and

electrical

components could be damaged.

Sieving aids

Stretching the sieve weave due to
excessive charge weight.

The precision of the analytical
sieve can be impaired.

Sieved product
distribution

Use the long threaded rods when
working with only one to three
sieves.

Overly long rods, extending
above the stack, will interfere
with distribution of the product
due to their intrinsic resonance.

Wet sieving

Never place the AS200 directly in
a basin for operation

Fatal electrical shock hazard.

Line power socket not protected
with a ground fault interrupter.

Fatal electrical shock hazard.

Water volume too great.

Water backing up in the sieve
stack can cause overloading
and thus may damage or
destroy the sieve weave.

Cleaning

Never clean the AS200 with
running water.

Fatal electrical shock hazard.

Advertising