Reusing the shipping nuts, Electrical connection – Retsch AS 200 User Manual

Page 8

Advertising
background image

31.07.2012

Retsch GmbH

8

SM

Set up the AS200 only on a stable laboratory bench to avoid un-
pleasant transmission of vibrations.

*

Loosen the hexagonal nuts SM using an open-ended

spanner until the sieve plate moves freely

Unscrew the wing screws A (shipping nuts) with hexagonal

nuts.

Mount the appropriate sieve clamping device

(See the

chapter on operation for instructions)

Keep the wing screws A with hexagonal nuts (shipping nuts)

in a safe place for later transportation.

Reusing the shipping nuts

Screw the hexagonal nuts fully onto the wing screws

Screw the wing screws through the sieve plate into the

housing

Twist the hexagonal nuts downwards and tighten until the

sieve plate can no longer move

Operation with these shipping nuts in place, or shipping without
the these nuts installed, could cause damage to mechanical
components.

Electrical connection

*

Refer to the data plate for information on the voltage and
frequency at which the AS200 is to be operated.

*

Ensure that the values shown there correspond to those for
the local power supply.

*

Use the supplied power cord to connect the AS200 to the
power supply.

*

When connecting the power cord to the power source, be sure
that the supply circuit is fused in accordance with local codes.

Failure to observe the values on the data plate can cause
damage to either the electrical or the mechanical com-
ponents or both.

Advertising