9

Emissions

Characteristic noise values :

Noise measurement according to DIN 45635-031-01-KL3
The characteristic noise values are determined primarily by the set
amplitude width or sieve plate acceleration, the number of sieves
deployed and the nature of the material to be screened.

Example :

Sound power level L

WA

= 64 dB(A)

Workplace-related emission value L

pAeq

= 58.4 dB(A)

Operating conditions :

Material to be screened = quartz sand, grain size <1mm, 5 sieves
speed = 150 rpm

Maximum loading

Max. quantity of material to be screened = 5 kg,
max. sieve stack mass = 15 kg.
It should be noted that, despite an integrated counterweight, work
cannot be performed at 300 r.p.m. in every case. The large imbal-
ance of, for example, 6x 400-type sieves, can cause the AS400 to
move on the laboratory bench.
At a maximum sieve stack mass of >10 kg, we recommend that
200 r.p.m. should not be exceeded.

Protection systems

IP54 or IP20 around the sieve carrier passage

Equipment dimensions

Height: up to approx. 850 mm, Width: 540 mm, Depth : 507 mm
Weight:

approx. 73 kg without sieve stack and

without sieve clamping device

Base area required

540 mm x 507 mm; no safety clearances required.

The AS400 control can cause major imbalances due to movements of

the material being screened. It must therefore be installed on a suit-

able bench in the laboratory. The net weight of the AS400 control is 73

kg.