RKI Instruments 35-3000RKA-LEL User Manual

Page 17

Advertising
background image

Sample Draw Combustible Gas Detector • 17

5. Connect the tubing from the gas collection bag to the fixed flow regulator, then open

the clamp.

6. Screw the fixed flow regulator onto the zero air calibration cylinder. The gas collection

bag begins to fill.

7. When the bag is full, unscrew the fixed flow regulator from the cylinder.

8. Close the clamp, then disconnect the sample tubing from the fixed flow regulator.

Setting the Zero Reading

1. Open the clamp, then connect the sample tubing from the gas collection bag to the

sample draw detector’s inlet line. This step is not necessary if you verified a fresh air
environment earlier in this procedure.

2. Allow the reading to stabilize for approximately 1 minute.
3. Verify a voltmeter reading of 100 mV (

±

2 mV).

4. If necessary, use a small flat-blade screwdriver to adjust the zero potentiometer until

the voltmeter reading is 100 mV (

±

2 mV).

Steps 5 and 6 are not necessary if you verified a fresh air environment earlier in this
procedure.

5. Allow the sample draw pump to draw out any residual gas in the gas collection bag.

6. Disconnect the sample tubing from the inlet line, then close the clamp.

Setting the Response Reading

1. Connect the sample tubing from the gas collection bag to the fixed flow regulator,

then open the clamp.

2. Screw the fixed flow regulator onto the calibration gas cylinder. The gas collection bag

begins to fill.

3. Unscrew the fixed flow regulator from the cylinder when the gas collection bag

appears full.

4. Close the clamp, then disconnect the sample tubing from the fixed flow regulator.

5. Open the clamp, then connect the sample tubing from the gas collection bag to the

inlet line at or near the sample draw detector’s INLET fitting.

6. Allow the sample draw detector to respond to the calibrating sample for

approximately 1 minute.

7. When the reading on the voltmeter stabilizes, verify that the reading matches the

response reading (

±

2 mV) you determined earlier.

8. If necessary, use the span potentiometer on the amplifier to adjust the reading to

match the correct response reading.

9. Allow the sample draw pump to draw out any residual gas in the gas collection bag.

10. Disconnect the sample tubing from the sample draw detector’s inlet line, then close

the clamp.

11. Reconnect the incoming sample line.

12. Wait 1 to 2 minutes to allow the combustible gas reading to decrease and stabilize.

Advertising