Lel sam pl e draw housing, Sample draw combustible gas detector • 9, Transmitte r, in terna lly wired – RKI Instruments 35-3000RKA-LEL User Manual

Page 9

Advertising
background image

Sample Draw Combustible Gas Detector • 9

5. Route the cable or wires in conduit leading from the sample draw detector to the

recording or monitoring device and power.

6. Connect the cable shield to an available chassis ground at the device end.

Figure 4: Wiring the Sample Draw Combustible Gas Detector

4/20 (F eedback)

Gr ound

+

WHT GR N BLK GN D

2 4V

+

_

L E L/ O2

AM P

P- AM P

4 / 20

GREEN

BLAC K

W HI TE

RED

PUMP ASSY
INTERNALLY
WIRED

24VD C

GN D

24V

4/20

RD

PUM P

PSW

TP

+

TP

ZERO

SPAN

GND

24V

4-20

RED

WHT

GRN

BLK

POWER/SIG

SENSOR

P/N 57-1050RK

REV. 0

NOT
USED
ON THIS
VE RS ION

H

N

1 15VAC

CG ND

PRES SURE SWITCH
INTERNALLY WIRED.

LEL Sam pl e Draw Housing

LEL DETECTO R IN
S AMPLE CHAMBER
(INT ERNALLY WIRED)

PCB IN SINGLE POINT
S AMPLE DRAWING
DETECTOR ASSY

Transmitte r, In terna lly Wired

Advertising