26 • 35-3000RK-10 PPM Hexane Sample Draw Detector

8.

If necessary, use the span pot on the amplifier to adjust the reading to match the
correct mV output.

9.

Allow the sample draw detector to draw out any residual gas in the gas collection
bag.

10. Disconnect the sample tubing from the sample-draw detector’s inlet sample line.

11. Remove the test jacks from the test points.

12. Close the housing door.

13. Unscrew the dispensing valve from the calibration cylinder.

14. Wait 1 to 2 minutes to allow the combustible gas reading to decrease.

15. Store the components of the calibration kit in a safe and convenient place.

Parts List

Table 2 lists replacement parts and accessories for the sample draw detector.

Table 2: Parts List

Part Number

Description

17-2593RK

Brass insert (for INLET and EXHAUST fittings)

17-2683RK

Front ferrule (for INLET and EXHAUST fittings)

17-2688RK

Back ferrule (for INLET and EXHAUST fittings)

30-0610RK

Pump

33-0163RK

Filter (Boston DFU9933-05-DQ)

33-2001RK-01

6 inch humidifier tube for fixed system calibration kit

71-0151RK

35-3000RKA-10 Operator’s Manual (this document)

81-0023RK-21

Calibration cylinder (500 ppm hexane in air, 34 liter)

81-0083RK-23

Calibration cylinder (500 ppm hexane in air, 103 liter)

81-0076RK-01

Zero air calibration cylinder (34 liter)

81-0076RK-03

Zero air calibration cylinder (103 liter)

81-1001RK

Dispensing valve (with knob)

81-1054RK

Demand flow regulator

81-1126RK

Gas collection bag (2 liter)

SG-8521

MOS combustible detector for hexane