Description, Housing – RKI Instruments 35-3000RKA-10 User Manual

Page 5

Advertising
background image

35-3000RK-10 PPM Hexane Sample Draw Detector • 5

Description

This section describes the components of the sample draw detector. It is a 4 to 20 mA type
detector head. The sample draw detector consists of the housing, flow system, and
detection system.

Figure 1: Component Location

Housing

The sample draw detector’s fiberglass housing is weather- and corrosion-resistant. It is
suitable for installation where general purpose equipment is in use.

The housing door is hinged on the left side and is secured by two latches on the right side.
The flowmeter and status LEDs are visible through a window in the housing door.

Four mounting feet are attached to the back of the housing (one at each corner). Use the
mounting feet to install the housing to a vertical surface. Use the two conduit hubs on the
bottom of the housing to make wiring connections.

An aluminum subpanel is mounted to the interior of the housing. The sample draw
detector’s internal components are mounted to the subpanel.

Factory Use Jumper Strip

Amplifier Plate

Load Voltage Pot

3/4" Conduit Hub, 2X

Detector
Terminal Strip

Mounting Foot, 4X

Flow Block

ppm Hexane
Detector

Interconnect
Terminal Strip

Detector/Amplifier
Terminal Strip

1-2 Hex
3-4 H2

Gas Select

1

3
4

2

P WR / S IG

S

W

B

R

D E TE C TOR

SPAN

ZERO

Pilot LED

Flow Adjust Pot

Exhaust

Controller
Terminal Strip

Amplifier

- Test Point, Black

Gas Selection Jumper,

In Hexane Position

Factory Set Pots

Chamber
Not Used

Zero Pot

Span Pot

Inlet

Flow Alarm Setpoint
Adjustment Screw

Factory Use Jumper Strip

+ Test Point, Red

Hydrophobic Filter

(Behind Main

Circuit Board)

Main Circuit Board

Fail LED

Pump/Flow Switch

Terminal Strip

Bypass v alve

Flowmeter