35-3010RKA-08 Sample-Draw Detector • 7

Oxygen Transmitter

The oxygen transmitter is mounted to the right of the LEL transmitter and above the main
circuit board. The amplifier includes the amplifier type selector, two internally wired
terminal strips, span pot, zero pot, and test points.

Figure 4: Oxgyen Transmitter Component Location

Transmitter Type Selector

The transmitter type selector is near the bottom left corner of the amplifier. It is to the left
of the detector terminal strip and below the span pot.

The transmitter included with the sample-draw detector is designed for use with RKI’s
oxygen and toxic gas sensors. The transmitter type selector determines for which sensor
the amplifier is intended. For oxygen transmitters, a jumper block is installed over the
OXY

selector.

Detector Terminal Strip

The detector terminal strip is the four-point terminal strip near the bottom edge of the
transmitter. It is factory wired to the transmitter.

Interconnect Terminal Strip

The interconnect terminal strip is the four-point terminal strip near the top edge of the
transmitter. It is factory wired to the main circuit board.

Span/Zero Pots

The span and zero pots are located on the left side of the transmitter and are used for
calibration. Use the span pot to make adjustments to gas response readings and the zero
pot to make adjustments to the zero reading

Test Points

The test points are located on the top of the transmitter on either side of the transmitter
interconnect terminal strip. They are labeled TP+ and TP-. A 100 mV - 500 mV output is
available at these test points for use during calibration.

TP

-

TP

+

4/20

FB

+

+

BATT

24V

Sensor Terminal
Strip

BK

W

OXY

SPAN

TOXIC

RD

TOXIC

OXY

G

Interconnect
Terminal Strip

Test Point TP-

Transmitter

Type Selector,

Set To OXY

Span Pot

ZERO

Test Point TP+

Zero Pot