RKI Instruments 35-3010RKA-08 User Manual

Page 9

Advertising
background image

35-3010RKA-08 Sample-Draw Detector • 5

Flow Blocks

Both flow blocks are located in the lower right corner of the sample-draw detector. The
oxygen sensor is installed in the larger flow block. The smaller flow block houses the IR
CH

4

sensor. The flow blocks route the sampled air to each sensor.

EXHAUST Fitting

The EXHAUST fitting on the bottom of the housing allows the gas sample to exit the
sample-draw detector. The EXHAUST fitting accepts 1/4 in. rigid tubing. See
“Installation” on page 10 to connect tubing to the EXHAUST fitting.

Detection System

The detection system consists of the gas sensors, LEL and oxygen transmitters, preamp
circuit board, and the main circuit board.

IR CH

4

Sensor

The infrared CH

4

sensor is installed in the smaller flow block in the lower right corner. A

small circuit board with a cable mates to it and retains it in the block. The cable is wired to
the main PCB.

Oxygen Sensor

The oxygen sensor is installed in the lower right of the larger flow block. It consists of a
cylindrical body which houses the detection elements and a cable which terminates in a
round 7-pin male connector. The sensor connector mates to a 7-position socket that is
wired to the main PCB.

Carbon Monoxide Sensor

The CO sensor position in the flow block is occupied by a plastic dummy plug in the
35-3010RKA-08.

Hydrogen Sulfide Gas Sensor

The H

2

S sensor position in the flow block is occupied by a plastic dummy plug in the

35-3010RKA-08.

Advertising