10 • 35-3010RKA-08 Sample-Draw Detector

Installation

This section describes procedures to mount the sample-draw gas detector in the
monitoring environment and wire the sample-draw detector to power and an external
device.

Mounting the Sample-Draw Detector

1.

Select the mounting site. Consider the following when you select the mounting site.

Is there enough room to open the housing door and make wiring connections at
the bottom of the housing and tubing connections at the right of the housing?

Make sure there is sufficient room to perform start-up, maintenance, and
calibration procedures.

Figure 6: Outline and Mounting Dimensions

7.

Close and latch the housing door.

NOTE:

The sample-draw detector is shipped with the mounting feet “tucked under” the
housing to protect the mounting feet during shipment.

8.

Slightly loosen the screw that secures one of the mounting feet to the housing, then
rotate the mounting foot 180 degrees (see Figure 6).

Ø .31 x .50 slot, 4X

8.31

.18

Tube Fitting for 1/4" OD

Tube, 2X

3/4" Conduit Hub, 2X

12.55

Door Latch

10.00

14.94

14.55

15.44