Troubleshooting – RKI Instruments 65-2330RK User Manual

Page 14

Advertising
background image

10 • 65-2330RK H

2

S Transmitter

7.

Reinstall the junction box cover.

8.

When the controller display reading falls below the alarm setpoints, return the
controller to normal operation.

9.

Store the components of the calibration kit in a safe place.

Quarterly

Calibrate the H

2

S transmitter as described in “Calibration” on page 13. See the calibration

frequency discussion in “Calibration Frequency” on page 13 to determine if a quarterly
calibration schedule fits your needs.

Troubleshooting

The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the H

2

S transmitter.

NOTE:

This troubleshooting guide describes transmitter problems only. See the
controller operator’s manual for problems you may encounter with the
controller.

Table 2: Troubleshooting H

2

S Transmitter

Condition

Symptom(s)

Probable Causes

Recommended Action

Fail Condition

• Controller indicates a

fail condition.

• The transmitter wiring

to the controller is
disconnected or
misconnected.

• The plug-in sensor is

not properly plugged
into the sockets in the
detector housing body.

• The wiring from the

detector to the
amplifier is
disconnected or
misconnected.

• The transmitter’s zero

reading is low enough
to cause a fail
condition.

• The transmitter is

malfunctioning.

1. Verify that the transmitter wiring to the

controller is correct and secure.

2. Confirm that the plug-in sensor is

installed properly.

3. Verify that the wiring from the detector

to the amplifier is correct and secure.

4. Perform a fresh air adjustment. A full

calibration is recommended.

5. If the fail condition continues, replace

the plug-in sensor as described later
in this section.

6. If the fail condition continues, contact

RKI for further instruction.

Advertising