RKI Instruments 65-2330RK User Manual

Page 18

Advertising
background image

14 • 65-2330RK H

2

S Transmitter

Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator.

Preparing for Calibration

NOTE:

Calibrating the H

2

S transmitter may cause alarms. Be sure to put the controller

into its calibration program or disable external alarms before calibration.

WARNING: Do not remove the detector housing cap or junction box cover while the

circuits are energized unless the area is determined to be non-hazardous.
Keep the detector housing cap and junction box cover tightly closed during
operation.

1.

Unscrew and remove the junction box cover.

2.

Set a voltmeter to measure in the millivolt (mV) range.

3.

Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead
into the red (+) test point; plug the negative lead into the black (-) test point.

4.

Use the following formula to determine the correct test points output for the
calibrating sample.

Output (mV) = (calibrating sample/fullscale) X 400 + 100

For example, with a calibrating sample of 25 PPM H

2

S and a fullscale setting of

100 PPM, the correct output is 200 mV.

200 (mV) = (25/100) X 400 +100

5.

Screw the calibration cup onto the detector housing.

Setting the Zero Reading

NOTE

If you can verify that the H

2

S transmitter is in a fresh air environment, you do not

need to apply zero air to the detector before adjusting the zero reading.

1.

Screw the regulator into the zero air calibration cylinder.

2.

Use the sample tubing to connect the regulator to the calibration cup.

3.

Turn the regulator knob counterclockwise to open the regulator.

4.

Allow the gas to flow for two minutes, then verify a reading of 100 mV (± 2mV). If
necessary, use the zero pot on the amplifier to adjust the reading to 100 mV (± 2mV).

5.

Turn the regulator knob clockwise to close the regulator.

6.

Unscrew the regulator from the zero air calibration cylinder.

7.

Leave the sample tubing connected to the regulator and the calibration cup.

Setting the Response Reading

1.

Screw the regulator into the calibration cylinder. Verify that the calibration gas is
representative of the transmitter’s target gas.

2.

Turn the regulator knob counterclockwise to open the regulator.

3.

Allow the gas to flow for two minutes, then verify that the reading matches the
response reading (± 2mV) you determined earlier. If necessary, use the span pot on the
amplifier to adjust the reading to match the correct response reading.

Advertising