Replacing components of the h, S transmitter – RKI Instruments 65-2330RK User Manual

Page 15

Advertising
background image

65-2330RK H

2

S Transmitter • 11

Replacing Components of the H

2

S Transmitter

This section includes a procedure to replace the H

2

S plug-in sensor, the entire detector

assembly, the hydrophobic membrane, and the amplifier. In most cases, it is not necessary
to replace the entire detector assembly.

Replacing the Plug-In H

2

S Sensor

CAUTION: The sensor contains electrolyte which is a dilute acid. Do not disassemble the sensor

when replacing it with a new one. If sensor electrolyte comes in contact with your
skin, wash affected area thoroughly with soap and water.

1.

Turn off the controller.

2.

Turn off or unplug power to the controller.

3.

Unscrew the detector housing cap from the detector housing body. Make sure not to
lose the cap gasket.

4.

Unplug and remove the H

2

S sensor.

5.

Carefully plug the replacement sensor into the four-socket pattern that is located in
the detector housing.

6.

Make sure the cap gasket is in place and screw the detector housing cap back onto the
detector housing body.

7.

Turn on or plug in power to the controller.

8.

Turn on the controller and place into normal operation.

CAUTION:

Allow the replacement sensor to warm up for 5 minutes before you continue with the
next step.

Slow or No
Response/
Difficult or
Unable to
Calibrate

• Transmitter responds

slowly or does not
respond to response
test.

• Unable to accurately

set the zero or
response reading
during calibration.

• Transmitter requires

frequent calibration.

Note: Under “normal”
circumstances, the
transmitter requires
calibration once every 3
months.

Some applications

may require a more
frequent calibration
schedule.

• The calibration cylinder

is low, out-dated, or
defective.

• The hydrophobic

membrane in the
detector housing cap is
wet or clogged with dirt
or other particulates.

• The calibration gas is

not an appropriate
concentration.

• The transmitter is

malfunctioning.

1. Verify that the calibration cylinder

contains an adequate supply of a
fresh test sample.

2. Check the face of the detector

housing cap and remove any
particulate contamination from the
hydrophobic membrane if necessary.
If the membrane appears saturated
with contamination or damaged,
replace the membrane as described
in the next section.

3. Verify that the calibration gas

concentration is appropriate for the
transmitter. Zero emission air (20.9%
oxygen) is normally used for a zero
adjustment if the environment is
suspect and 25 PPM H

2

S in nitrogen

is normally used for a response test.

4. If the calibration/response difficulties

continue, replace the plug-in sensor
as described later in this section.

5. If the calibration/response difficulties

continue, contact RKI for further
instruction.

Condition

Symptom(s)

Probable Causes

Recommended Action

Advertising