Installation – RKI Instruments 65-2330RK User Manual

Page 9

Advertising
background image

65-2330RK H

2

S Transmitter • 5

calibration cup on the detector during calibration.

Installation

This section describes procedures to mount the H

2

S transmitter in the monitoring

environment and wire the transmitter to a controller.

Mounting the H

2

S Transmitter

1.

Select a mounting site that is representative of the monitoring environment. Consider
the following when you select the mounting site.

Select a site where the transmitter is not likely to be bumped or disturbed. Make
sure there is sufficient room to perform start-up, maintenance, and calibration
procedures.

Select a site that is at normal breathing level.

Figure 3: Mounting the H

2

S Transmitter

2.

At the monitoring site you select, hang or mount the junction box with the detector
facing down (see Figure 3).

.38

.75

3.10

3/4 NPT
Female

Rubber
Spacers
3X

3.65

H2S Detector

J-Box

1 1/2-20 Thread for
Calibration Cup

6.8 Max