Calibration – RKI Instruments 65-2340RKSS User Manual

Page 18

Advertising
background image

14 • 65-2340RKSS Toxic Gas Transmitter

Calibration

This section describes how to calibrate the toxic gas transmitter. It includes procedures to
prepare for calibration, set the zero (fresh air) reading, set the span (response) reading,
and return to normal operation. It describes calibration using a calibration kit that
includes a calibration cup, calibration gas, sample tubing, and a fixed flow regulator with
an on/off knob. The required sample flow rate and calibration cup depend on the
transmitter being calibrated. Table 3 below lists the required calibration cups and
regulators for the various toxic gas transmitters.

WARNING: Not using the recommended calibration cup, sample flowrate, and specified

flow direction (for 81-1138RK calibration cup only) will result in an
inaccurate calibration. Make sure to use the calibration cup and sample
flow rate listed in Table 3 for your target gas when performing a
calibration. Also make sure to connect the sample tubing to the calibration
cup port that results in the specified flow direction for the 81-1138RK
calibration cup.

Preparing for Calibration

1.

Place the controller into its calibration mode or disable external alarms.

2.

Unscrew and remove the junction box cover.

3.

Set a voltmeter to measure in the millivolt (mV) range.

4.

Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead
into the red + test point; plug the negative lead into the black - test point.

5.

Use the following formula to determine the correct test points output for the
calibrating sample.

Output (mV) = (calibrating sample/fullscale) X 400 + 100

Table 3: Required Calibration Cups and Flow Rates

Transmitter Type

Required

Calibration Cup

Required Regulator/

Flowrate for Calibration

Gas

Flow Direction

Specified on

Calibration Cup

Arsine (AsH

3

)

81-1138RK

81-1051RK, 0.5 LPM (liters
per minute)

Yes

Chlorine (Cl

2

), 0 - 3.00 ppm

and 0 - 10.0 ppm

81-1138RK-CL2

81-1051RK, 0.5 LPM

No

Hydrogen Cyanide (HCN)

81-1138RK-HCN

81-1051RK, 0.5 LPM

Yes

Ammonia (NH

3

)

81-1138RK-NH3

81-1051RK-25, 0.25 LPM

No

Phosphine (PH

3

)

81-1138RK

81-1051RK, 0.5 LPM

Yes

Sulphur Dioxide (SO

2

)

81-1138RK

81-1051RK, 0.5 LPM

Yes

Advertising