65-2340RKSS Toxic Gas Transmitter • 1

Overview

This manual describes the 65-2340RKSS toxic gas transmitter. This manual also describes
how to install, start up, maintain, and calibrate the toxic gas transmitter when used with a
gas monitoring controller. A parts list at the end of this manual lists replacement parts and
accessories for the toxic gas transmitter.

The 65-2340RKSS toxic gas transmitter can be used for various target gases which are
listed in Table 1 below. For a particular target gas, the part number of the toxic transmitter
will be of the format 65-2340RKSS-XXX. For example, 65-2340RKSS-NH3 is for ammonia
(NH

3

) detection.

Specifications

Table 1 lists specifications for the toxic gas transmitter.

WARNING: When using the 65-2340RKSS, you must follow the instructions and

warnings in this manual to assure proper and safe operation of the
65-2340RKSS and to minimize the risk of personal injury. Be sure to
maintain and periodically calibrate the 65-2340RKSS as described in this
manual.

Table 1: Specifications

Target Gas

65-2340RKSS-ASH3: Arsine (AsH

3

)

65-2340RKSS-CL2: Chlorine (Cl

2

)

65-2340RKSS-CL-10: Chlorine (Cl

2

)

65-2340RKSS-HCN: Hydrogen Cyanide (HCN)
65-2340RKSS-NH3: Ammonia (NH

3

)

65-2340RKSS-SO2: Sulphur Dioxide (SO

2

)

65-2340RKSS-PH3: Phosphine (PH

3

)

Sampling Method

Diffusion

Detection Range

AsH

3

: 0 - 1.50 ppm

Cl

2

: 0 - 3.00 ppm

Cl

2

: 0 - 10.0 ppm

HCN: 0 - 15.0 ppm
NH

3

: 0 - 75.0 ppm

SO

2

: 0 - 6.00 ppm

PH

3

: 0 - 1.00 ppm

Accuracy

± 10% of reading or ± 5% of full scale (whichever is greater)

Signal Output

4 to 20 mA

Response Time

• Cl

2

and NH

3

: T90 in 90 seconds

• All others: T90 in 45 seconds

Operating Temperature &

Humidity

• 14°F to 104°F (-10°C to 40°C)
• 20% to 90% Relative Humidity