16 • 65-2397RK CO

2

Transmitter

Calibration

This section describes how to calibrate the CO

2

transmitter. It includes procedures to

prepare for calibration, set the zero reading, set the response reading, and return to
normal operation.

WARNING: The controller is not an active gas monitoring device during the calibration

procedure.

NOTE:

The following procedure assumes the use of a calibration kit which includes a
calibration gas cylinder, a 100% nitrogen cylinder, a 0.5 LPM fixed flow regulator
with an on/off knob, a calibration cup for the detector, and a short piece of
sample tubing to connect the regulator to the calibration cup.

Preparing for Calibration

1.

Unscrew and remove the junction box cover.

2.

Set a voltmeter to measure in the millivolt (mV) range.

3.

Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead
into the red (+) test point; plug the negative lead into the black (-) test point.

4.

Use the following formula to determine the correct test points output for the
calibrating sample.

Output (mV) = (calibrating sample/fullscale) X 400 + 100

For example, with a calibrating sample of 2.5 %Vol. and a fullscale setting of
5 %Vol., the correct output is 300 mV.

300(mV) = (50/100) X 400 +100

5.

Screw the calibration cup onto the detector housing.

6.

Place the controller into its calibration mode or disable external alarms.

NOTE:

Calibrating the CO

2

transmitter may cause alarms. Be sure to put the controller

into its calibration mode or disable external alarms before continuing.

Setting the Zero Reading

Since there is a background of CO

2

in air of typically 300 - 600 ppm (0.03 - 0.06 %volume),

it is necessary to use a calibration kit with a 100% nitrogen cylinder to set the zero signal of
a CO

2

transmitter with a low range detector. Fresh air can be used to zero the transmitter

if a 0-50 %volume or 0-100 %volume detector is being used.

1.

Screw the regulator into the 100% nitrogen calibration cylinder.

2.

Use the calibration kit sample tubing to connect the regulator to the calibration cup.

3.

Turn the regulator’s on/off knob counterclockwise to open the regulator.

4.

Allow the gas to flow for one minute and verify a reading of 100 mV (±2 mV). If
necessary, use the zero pot on the amplifier to adjust the reading to 100 mV (±2 mV).

5.

Turn the regulator’s on/off knob clockwise to close the regulator.