Installation – RKI Instruments 65-2397RK User Manual

Page 9

Advertising
background image

65-2397RK CO

2

Transmitter • 5

Installation

This section describes procedures to mount the CO

2

transmitter in the monitoring

environment and wire the transmitter to a controller.

Mounting the CO

2

Transmitter

1.

Select a mounting site that is representative of the monitoring environment. Consider
the following when you select the mounting site.

Select a site where the transmitter is not likely to be bumped or disturbed. Make
sure there is sufficient room to perform start-up, maintenance, and calibration
procedures.

Mount the transmitter at least six feet from the floor to minimize the possibility of
someone breathing on the unit. The exhaled CO

2

may cause an alarm.

Figure 3: Mounting the CO

2

Transmitter

2.

At the monitoring site you select, hang or mount the junction box with the detector
facing down (see Figure 3).

Rubber
Spacers,
3X

3.10

.75

.38

3.65

3/4 NPT

Female

1 1/2-20 Thread for

Calibration Cup

J-Box

IR CO2

Detector

6.80 MAX