Troubleshooting – RKI Instruments 65-2405RK-04 User Manual

Page 17

Advertising
background image

65-2405RK-04 Combustible Gas Transmitter • 13

Troubleshooting

The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the combustible gas transmitter.

NOTE:

This troubleshooting guide describes transmitter problems only. See the
controller operator’s manual for problems you may encounter with the
controller.

Replacing Components of the Combustible Gas Transmitter

This section includes procedures to replace the combustible gas detector and amplifier.

Replacing the Combustible Gas Detector

NOTE:

The combustible gas detector is installed on the junction box that does not house
the amplifier. This junction box was likely installed somewhere difficult to
access.

1.

Turn off the controller.

2.

Turn off or unplug power to the controller.

3.

Unscrew the calibration cup/splash guard from the detector.

Table 2:Troubleshooting the Combustible Gas Transmitter

Condition

Symptom(s)

Probable Causes

Recommended Action

Fail Condition

• Controller indicates a

fail condition.

• The detector-to-

amplifier or amplifier-
to-controller wiring is
disconnected or
misconnected.

• The transmitter’s zero

reading is low enough
to cause a fail
condition.

• The transmitter is

malfunctioning.

1. Verify that the detector-to-amplifier

and amplifier-to-controller wiring is
correct and secure.

2. Calibrate the transmitter.
3. If the fail condition continues, replace

the detector.

4. If the fail condition continues, contact

RKI for further instruction.

Slow or No
Response/
Difficult or
Unable to
Calibrate

• Transmitter responds

slowly or does not
respond to response
test.

• Unable to accurately

set the zero or
response reading
during calibration.

• Transmitter requires

frequent calibration.

Note: Under “normal”
circumstances, the
transmitter requires
calibration once every 3
months.

Some applications

may require a more
frequent calibration
schedule.

• The calibration cylinder

is low, out-dated, or
defective.

• The calibration gas

flow rate is too low.

• The calibration gas is

not an appropriate
concentration.

• The transmitter is

malfunctioning.

• The flexible tubing to

the calibration cup/
splash guard has
become kinked,
blocked, or
disconnected.

1. Verify that the calibration cylinder

contains an adequate supply of a
fresh test sample.

2. Verify that the regulator used for

calibration is a 0.5 LPM regulator.

3. Verify that the calibration gas

concentration is appropriate for the
transmitter. The concentration should
be in the detection range, preferably
about half of the detection range.

4. Verify that the flexible tubing to the

calibration cup/splash guard is clear
and connected to the calibration cup/
splash guard.

5. If the calibration/response difficulties

continue, replace the detector.

6. If the calibration/response difficulties

continue, contact RKI for further
instruction.

Advertising