Calibration frequency – RKI Instruments 65-2405RK-04 User Manual

Page 19

Advertising
background image

65-2405RK-04 Combustible Gas Transmitter • 15

amplifier as shown in Figure 4 on page 9 of this manual. If controller leads or detector
leads were removed during this procedure, refer to Table 3 and Table 4 below.

NOTE:

When a transmitter is first powered up with a new amplifier, the initial output
may be either high or below zero depending on the setting of the zero pot. Be
sure to make arrangements so that this does not cause unwanted alarms.

10. Turn on power to the controller.

11. Turn on the controller and place it into normal operation.

NOTE:

When first powered up, the transmitter will enter about a one minute period
when the 4-20 mA output is stabilizing and may be above the controller alarm
points or well below zero momentarily. RKI controllers have a one minute
warmup period when the controller does not display any gas reading or give any
alarm indication. The combustible gas transmitter’s 4-20 mA signal should be
stable by the time the controller’s warmup period is over.

12. Allow the transmitter to warm-up for 5 minutes.

13. Calibrate the transmitter as described in “Calibration” on page 17 of this manual.

Calibration Frequency

Although there is no particular calibration frequency that is correct for all applications, a
calibration frequency of every 3 to 6 months is adequate for most combustible gas
transmitter applications. Unless experience in a particular application dictates otherwise,
RKI Instruments, Inc. recommends a calibration frequency of every 3 months.

Table 3:Reconnecting the Amplifier

to the Controller

Amplifier Controller

Terminal Strip

Controller

Transmitter Terminal

Strip (typical)

PWR/SIG “-”

- (DC -)

PWR/SIG “S”

S (4 - 20 mA In)

PWR/SIG “+”

+ 24V

Table 4:Reconnecting the LEL

Detector to the Amplifier

Amplifier Detector

Terminal Strip

Detector Lead

DETECTOR “R”

RED

DETECTOR “W”

WHT

DETECTOR “G”

GREEN

DETECTOR “B”

BLK

Advertising