Start up – RKI Instruments 65-2432RK User Manual

Page 10

Advertising
background image

65-2432RK CO Transmitter • 10

Start Up

This section describes procedures to start up the CO transmitter and place the
transmitter into normal operation.

Introducing Incoming Power

1.

Complete the installation procedures described earlier in this manual.

2.

Verify that the power wiring to the controller is correct and secure. Refer to the
controller instruction manual.

3.

Turn on or plug in the incoming power at the power source end, then turn on the
controller.

4.

Verify that the controller is on and operating properly. Refer to the controller
instruction manual.

CAUTION:

Allow the transmitter to warm up for 5 minutes before you continue with the next

section, “Setting the Zero Signal.”

Setting the Zero Signal

CAUTION:

If you suspect the presence of the target gas in the monitoring environment, use the
calibration kit and the zero air calibration cylinder to introduce “fresh air” to the

sensor and verify an accurate zero setting.

1.

Verify that the transmitter is in a fresh air environment (environment known to be
free of carbon monoxide, other toxic and combustible gases and of normal oxygen
content, 20.9%).

2.

Unscrew and remove the junction box cover from the junction box.

3.

Set a voltmeter to measure in the millivolt (mV) range.

4.

Plug the voltmeter leads into the test points on the amplifier. Plug the positive
lead into the test point labeled

TP+

; plug the negative lead into the test point

labeled

TP-

.

5.

Verify a voltmeter reading of 100 mV (±2 mV).

6.

If necessary, use a flat-blade screwdriver to adjust the zero potentiometer until the
voltmeter reading is 100 mV (±2 mV).

7.

Remove the voltmeter leads and secure the junction box cover to the junction box.

Advertising