RKI Instruments 65-2432RK User Manual

Page 15

Advertising
background image

65-2432RK CO Transmitter • 15

11. Reconnect the detector leads to the detector terminal strip as shown in Table 3 and

Figure 3, Wiring the CO Transmitter to a Controller.

12. Reinstall the junction box cover.

13. Turn on or plug in incoming power at the power source end.

14. Turn on the controller.

CAUTION:

Allow the sensor to warm up for 5 minutes before you continue with the next step.

15. Calibrate the CO transmitter as described in the Calibration section of this manual.

Replacing the CO detector

NOTE:

In most cases, it is only necessary to replace the CO sensor.

1.

Turn off the controller.

2.

Turn off or unplug incoming power at the power source end.

3.

Remove the junction box cover.

4.

Disconnect the detector leads from the detector terminal strip. Note the position of
the color-coded leads as you remove them.

5.

Unscrew the detector from the junction box.

6.

Guide the detector leads of the replacement detector through the bottom conduit
hub of the junction box, then screw the mounting threads of the detector into the
conduit hub.

7.

Connect the detector leads to the detector terminal strip as shown in Table 4 and
Figure 3, Wiring the CO Transmitter to a Controller.

8.

Reinstall the junction box cover.

9.

Turn on or plug in incoming power at the power source end.

10. Turn on the controller.

Table 3: Reconnecting the CO Detector to the Amplifier

CO Detector Lead

Amplifier

Interconnect Terminal

Strip

Black

TOXIC BK

Red

TOXIC RD

Table 4: Connecting the Replacement CO Detector to the Amplifier

CO Detector Lead

Amplifier

Interconnect Terminal

Strip

Black

TOXIC BK

Red

TOXIC RD