65-2640RK-04 M2A Transmitter Operator’s Manual

22

NOTE:

The low power alarm cannot be cleared using the ENTER button.

When the voltage increases to 10.0 volts, the low power alarm is cleared and the M2A will
begin its warm-up sequence.

Responding to a Low Power Condition

1.

Determine and correct the cause of the low power condition.

2.

When the input power increases above 10.0 volts, the M2A will begin its warm-up
sequence.

3.

Verify that the M2A enters normal operation after its warm-up sequence. If necessary,
perform a fresh air adjustment. See “Adjusting the Fresh Air Reading” on page 16.