65-2640RK-04 M2A Transmitter Operator’s Manual

26

Chapter 6: Gas Type Mode

Overview

This chapter describes how to use Gas Type Mode to select the M2A’s gas type. The gas
type determines the target gas and detection range.

CAUTION:

The target gas is factory set and does not normally need to be changed. The
standard setting for the 65-2640RK-04 is methane, but it may be different
depending on the calibration.

Gas Type Mode includes a 5-minute time-out feature. If you do not press a button for 5
minutes, the M2A automatically returns to normal operation.

NOTE:

If the M2A enters normal operation because of a program time-out, it enters a
warm-up period just as it does when it is first turned on.

Selecting the Gas Type

1.

While in normal operation or during the warm-up sequence, press and hold the UP/
YES and DOWN/NO buttons for five seconds. Release them when the following
screen appears.

2.

If you want to exit Gas Type Mode, press and release the DOWN/NO button. The
display will indicate NO CHANGE and the M2A will return to normal operation.

If you want to continue in Gas Type Mode, press and release the UP/YES button. The
top display line will indicate GasType? and the bottom display line will indicate the
target gas choice.

3.

Use the UP/YES or DOWN/NO button to scroll through the choices of target gas.
Table 6 below lists the choices of target gas for the M2A.

Table 6: Gas Types

Gas Type

Choices

Detection Range

HC

0 - 100% LEL

iBUTANE

0 - 100% LEL

PROPANE

0 - 100% LEL

HEXANE

0 - 100% LEL

HYDROGEN

0 - 100% LEL

METHANE

0 - 100% Volume

Select

GasType?