5 • Model GD-70D4H-XX Sample Draw Transmitter

Figure 3: Outline & Mounting Dimensions

Connecting Sample Lines to the GD-70D4H-XX

1.

Connect the hydrophobic filter to the tubing stub on the inlet.

If the hydrophobic filter is not used, or if you install the hydrophobic filter in a
different location, replace the tubing stub with 1/4’’ x 1/8’’ tubing as described in
“Inlet Fitting Tubing Stub” on page 3.

2.

Connect a length of sample tubing to the other side of the hydrophobic filter and route
it to the sampling area. If the hydrophobic filter is not used, or if you install the
hydrophobic filter in a different location, route tubing from inlet fitting to sample
area. Be sure to keep the length to a minimum. Consult RKI Instruments, Inc. for
lengths of more than 20 feet.

CAUTION:

Avoid loops or slumps in the incoming sample line. To reduce response time, keep the
incoming sample line as short as possible.

3.

Install 1/4“ O.D. x 1/8” I.D. Teflon PTFE sample tubing to the gas out fitting. Route

NOTE: A ll D im enions

Shown i n Inches

Inlet F itting

8.22

2.51

Ex haust Fitti ng

3/4" C onduit

Hub, 2X

9.88

7.01

1.43

8.40

4.83

8.00 M AX

11.75

11.23

Tubing S tub

0.85 M AX

Ø .34, 4X