Model GD-K7D2 (24 VDC) Sample Draw Transmitter • 1

Overview

This manual describes the 24 VDC version of the Model GD-K7D2 sample draw 4 - 20 mA
transmitter detector head. This manual also describes how to install, start up, maintain,
and calibrate the GD-K7D2 when it is used with a gas monitoring controller. A parts list at
the end of this manual lists replacement parts and accessories for the GD-K7D2.

Specifications

Table 1 lists specifications for the 24 VDC version of the GD-K7D2.

WARNING: When using the GD-K7D2, you must follow the instructions and warnings

in this manual to assure proper and safe operation of the GD-K7D2 and to
minimize the risk of personal injury. Be sure to maintain and periodically
calibrate the GD-K7D2 as described in this manual.

Table 1: Specifications

Target Gas & Detector Range

Refer to the RKI Instruments Inc. List of Detectable Gasses

Area Classification

Indoor, non-hazardous locations

Sampling Method

Sample-draw

Input Power

24 VDC

Signal Output

4 to 20 mA

Response Time

90% in 60 seconds

Accuracy

± 10% of reading or ± 5% of full scale (whichever is greater)

Flow Rate

Approximately 0.5 liters/minute

Recommended Sample Tubing

4 mm I.D. x 6 mm O.D. Teflon PTFE

Recommended Sample Tubing Length

See Table 2 in the Installation section on p. 10