Installation bracket – RKI Instruments GD-K7D2 (24 VDC) User Manual

Page 9

Advertising
background image

Model GD-K7D2 (24 VDC) Sample Draw Transmitter • 5

output during start-up and calibration procedures.

Gain pot

The gain pot is below the zero pot. It is used to make coarse span adjustments. The gain
pot is factory set and for adjustment by a field service technician only

.

Another factory adjust pot is located below the gain pot. Do not adjust this pot.

Output check pins

Two output check pins [marked 1 (+) and 2 (-)] are below the pots. Use the output check
pins to read the signal output of the GD-K7D2 with a milliampmeter during the start-up
and calibration procedures.

Pump

The pump is behind the sensor. The pump pulls the test sample into the GD-K7D2. The
pump operates on AC voltage that is generated by the GD-K7D2 from the 24 VDC input
power.

Lithium Battery Pack

A 3.6 volt lithium battery pack is above the pump on the right side of the GD-K7D2. The
lithium battery pack maintains a bias voltage on the gas sensor when the GD-K7D2 is not
receiving incoming power, such as during shipment or storage. If the GD-K7D2 is off
power for an extended period and the lithium battery pack is dead, the gas sensor
operation will be affected when the GD-K7D2 is started up. See Storing the GD-K7D2 in
the Maintenance section for battery pack replacement recommendations.

Installation Bracket

The L-shaped installation bracket facilitates mounting and wiring of the GD-K7D2. The
three mounting holes on the back of the bracket are used to mount it to a vertical surface.
The installation bracket includes the socket, external wiring terminal strip, cable bushing,
and sample fittings.

Figure 4: Installation Bracket Component Location

Socket for
Main Board

External Wiring
Terminal Strip

Mounting Holes

Cable Bushing

Gas In

Sample Fitting

Gas Out

SampleFitting

Factory Installed
Jumper

*

*

Jumper is factory installed when
GD-K7D2 is used with a controller
that does not have a built in flow
alarm circuit and flow alarm
wiring terminals.

Advertising