PT2E-1140

⑤ Operate the pump and adjust flowmeter at 1.5l/min.

⑥ When the calibration gas is introduced into the sensor, the indication of the indicator

rises. If the indication is not equal to the concentration value of calibration gas in 2
minutes after start of introduction, press the control key to the UP switch▲ or
DOWN switch▼ to allow the indication to match to the calibration gas concentration.

⑦ After adjustment, press SET switch● to decide.

SPAN light is changed from flickering to lighting, indication value on LCD is changed

from flickering to lighting and indicating that SPAN adjustment is completed.

Gas sampling bag

Gas bag for exhaust.

⑧ Upon completion of the span adjustment, remove the sampling bag from the IN side of

the gas check adaptor and press the control key to MAINTENANCE switch●.

Then, proceed to the zero adjustment again.

If the reading cannot be adjusted to the calibration gas concentration, sensor life will

be over. Put off the power switch and contact with our authorized agent.

(It is recommendable to replace the sensor every one year). For sensor replacement,

refer to “4-8 Sensor Replacement Method.”

Confirm that the replacement sensor is the same type as described

in the sensor type seal sticked on the detector.

CAUTION


Make zero and span adjustments again after sensor replacement.