Measures for storage or long-time shutdown, Recommendable spare parts list – RKI Instruments SD-705EC User Manual

Page 28

Advertising
background image

PT2E-1140

4−10 Measures for Storage or Long-time Shutdown

(1) Store the sensor as attached to the detector in a place not exposed to dust and water

splash.

The warm up time will be longer if the power is not supplied to the sensor for a long time.

If the sensor is kept for more than one month without power supply, gas calibration may

be required. So, it is recommendable to supply the power always.

(2) Storage conditions

Temperature :

-10 ∼ +40℃

Humidity

30 ∼ 80%RH 以下

Environmental conditions:

Place without organic solvent and gas generation.

4−11 Recommendable Spare Parts List

No.

Description

Checking

interval

Replacement

interval

Q’ty/Detector

1

Sealing packing

1 year

5∼6 years

1 pce

2

Packing

1 year

5∼6 years

1 pce

3

Sensor

6 months

1 year

1 pce

4

Printer circuit

board(Display)

5∼6 years

1 set

5

Printed circuit board(Main)

7∼8 years

1 set

6

Fuse

8 years

1 pce

*1:The above replacement interval may vary depending on the conditions

of actual use. It does not mean the warranty period.

The replacement interval also may vary by the result of regular inspection.

*2 : The main reason of the printed circuit board is the deterioration of capacitors.


27