RKI Instruments SD-705EC User Manual

Page 32

Advertising
background image

PT2E-1140

③ Span adjustment impossible

Span adjustment impossible

Yes

Calibration gas prepared
correctly?

No

Use the correctly prepared gas for
re-calibration.

Yes

Yes

No

Carry out calibration according to
“4-3 Gas sensitivity calibration
method”.

Calibration gas supplied
correctly to the sensor?

No

Yes

Sensor deterioration.
Since the sensor must be replaced, contact
our local agent or RIKEN KEIKI.

There may be an
abnormality in the electric
circuit.
Turn OFF the power
supply to this unit and
contact our local agent or
RIKEN KEIKI.

The CD indication changes UP or DOWN
when the control key is brought nearer to
the UP▲ or DOWN▼ switch?

31