Operating procedures, Operating procedures (fixed temperature operation), How to start fixed temperature operation – Yamato Scientific DVS602 Constant Temperature Drying Oven User Manual

Page 19

Advertising
background image

17

4. Operating procedures

Operating procedures (fixed temperature operation)

How to start fixed
temperature
operation



MEASURED TEMP.

HEATER

ALARM

AUTO STOP

AUTO START

SET

TEMP.

OVER TEMP.

PROTECTOR

FIXED TEMP.

PROGRAM

1. Turn power ON. (Turn the ELB to ON.)

When power is turned ON, the initial values will be displayed for
about four seconds, then the initial screen will appear and the
current bath temperature, operation mode character and the
overheat prevention set temperature are displayed on each of the
displays.

MEASURED TEMP. display: Indicates the present bath temperature and various
characters.
Set temperature display: Indicates the set temperature and various characters.


Overheat prevention set temperature display: Indicates the set temperature for
the overheat prevention device.


For operation mode characters, see section page 11.

MEASURED TEMP.

HEATER

ALARM

AUTO STOP

AUTO START

ENTER

SET

TEMP.

OVER TEMP.

PROTECTOR

FIXED TEMP.

RUN

STOP

SUB

MENU

TIMER

FIXED
TEMP.

PROGRAM

PROGRAM



MEASURED TEMP.

HEATER

ALARM

AUTO STOP

AUTO START

ENTER

SET

TEMP.

OVER TEMP.

PROTECTOR

FIXED TEMP.

RUN

STOP

SUB

MENU

TIMER

FIXED
TEMP.

PROGRAM

PROGRAM



MEASURED TEMP.

HEATER

ALARM

AUTO STOP

AUTO START

ENTER

SET

TEMP.

OVER TEMP.

PROTECTOR

FIXED TEMP.

RUN

STOP

SUB

MENU

TIMER

FIXED
TEMP.

PROGRAM

PROGRAM

2. Selecting an operation mode

Press the FIXED

TEMP. key to display the fixed temperature

operation mode on the SET TEMP. display.

Display the character Fix

.







3. Setting the temperature

Press the FIXED TEMP. key again.

The character SV

that indicate a set temperature is

displayed on the MEASURED TEMP. display on which the current
set temperature flashes and the FIXED TEMP. lamp flashes.
Set a temperature using the ▼▲ keys.


4. Starting operation

Press the RUN/STOP key for about one second. Operation starts
and status of the FIXED TEMP. lamp changes from flashing to
illuminated.

5. Stopping operation

Press the RUN/STOP key for about one second. Operation stops,
the FIXED TEMP. lamp goes off and the screen switches to the initial
setting screen.


When you want to
correct setting
errors or confirm
settings

When you made a mistake during setting or reconfirm settings you
made, press the FIXED TEMP. key again to resume setting.
When you want to change set temperature during operation, press the

FIXED TEMP. key to enter the setting mode and change the set

temperature. After change has been made, press the ENTER key to
complete the process.

Advertising