4. Operating procedures

Preparing a program

MEASURED TEMP.

HEATER

ALARM

AUTO STOP

AUTO START

ENTER

SET

TEMP.

OVER TEMP.

PROTECTOR

FIXED TEMP.

RUN

STOP

SUB

MENU

TIMER

FIXED
TEMP.

PROGRAM

PROGRAM

② Select a program mode you want and press the ENTER key.

・When PrG1

is selected, End

appears on the

MEASURED TEMP. display and the number of steps
registered flashes on the SET TEMP. display.

・When PrG2

is selected, PAt

appears on the

MEASURED TEMP. display and the pattern number flashes
on the SET TEMP. display. To select a pattern select [1] or
[2] with the ▼▲ keys.

Pressing the ENTER key displays End on the MEASURED

TEMP. display and the number of steps registered flashes on
the SET TEMP. display.

・When PrG3

is selected, select [1], [2], or [3] following

the same procedures as for PrG2.


Since the sample program uses up to nine steps, input is allowed
in any of the program modes of PrG1, PrG2, or PrG3.

Here, how to register a program is explained using PrG3 as an

example.

MEASURED TEMP.

HEATER

ALARM

AUTO STOP

AUTO START

ENTER

SET

TEMP.

OVER TEMP.

PROTECTOR

FIXED TEMP.

RUN

STOP

SUB

MENU

TIMER

FIXED
TEMP.

PROGRAM

PROGRAM

4. Registering a program (inputting a program)

① Select PrG3 following the previous section 3.
② Input the number of program steps, step temperature, and step

time following the entries in the programming sheet.

③ Press the ENTER key. PAt number flashes. (End will

appear when PrG1 is selected. Proceed to section ⑥.)

④ Select an unused pattern from PAt1, PAt2, or PAt3 with the

▼▲ keys.

⑤ Press the ENTER key. End appears and the number of steps

“n” flashes.
*End is a character that indicates the total number of steps to
be used. In the example, you will input “9.”

⑥ Input the total number of program steps to be used “9” with the

▼▲ keys.

⑦ Press the ENTER key. The character Sv-1

that

indicates the set temperature of the first step is displayed and
the current set temperature flashes.

⑧ Set a temperature for the first step using the ▼▲ keys.